Priority Hardware Maintenance Agreements (HMA)

HMA Priority Services are offered in three levels to match the measuring envelope and array of sensors to be certified on your system.

Level 1 — covers bench-top video systems with one standard sensor.

Level 2 — covers mid-range floor model systems with up to two standard sensors or one advanced sensor.

Level 3 — covers large travel floor model systems with up to two standard sensors or one advanced sensor.

All three levels include two on-site service visits per year by an ISO 17025-2005 certified Field Service Technician, including:

  • As Found performance baseline testing and report
  • Cleaning and mechanical preventative maintenance of the ways and optics, and back-up of configuration files and system parameters.
  • Calibration of the optical system and X,Y,Z axes, using NIST Traceable calibration artifacts.
  • ISO 17025-2005 calibration report
  • As Left performance report, including calibration results and written report of any additional service recommendations

Additional sensors can be added to the HMA plan for a small additional fee. Why should you have an HMA?

 Standard Sensors Advanced Sensors
 Video
TP 20 or TP 200 touch probe
TTL Laser
DRS or DRX Laser
 TeleStar Laser
Scanning Probe
Feather Probe
Rainbow Probe

HMA Graph

Preventative Maintenance

Preventative Maintenance includes confirming proper operation of the system motion, optics, lighting and sensors, recommended cleaning and service for all mechanisms, and backing up configuration and parameter files.

Motion

  • Check, clean, lubricate and align lead screws / rod drives
  • Scale reader head alignment check / adjustment
  • Confirm motion acceleration and velocity within normal range

Optics

  • Camera Coaxial & Rotational alignment
  • Parfocalization
  • Video brightness and contrast
  • Zoom lens calibration
  • Field of view calibration

Lighting

  • Confirm normal operation

Electronics

  • Confirm power supply voltages
  • Check and clean all filters and fans

Sensors

  • Confirm proper operation of all sensors & deployment mechanisms

Documentation

  • Record as-found and as-left condition of system

Calibration

System calibration includes performing a baseline as-found accuracy test for the X, Y and Z-axes, and a calibration of each axis using the recommended artifact.

Baseline

  • X,Y Accuracy
  • X,Y Repeatability
  • X,Y Hysteresis
  • Z Accuracy
  • Z Repeatability
  • Z Hysteresis

Calibration

  • Z Cert
  • Z Verify
  • X,Y Cert
  • X,Y Verify

Calibration Report